Gold Member Since 2012
Audited Supplier
Guangzhou Maya Medical Equipment Co., Ltd.

Icu Patient Monitor, Patient Monitor, Patient Monitoring System manufacturer / supplier in China, offering My-C039 12 Lead ECG+NIBP+SpO2+Pr+Resp+Temp+Printer+Etco2+IBP ICU Patient Monitor, My-D017 500mA Medical X Ray Unit X-ray Machine, My-D014 Medical Supply 300mA Medical X-ray Machine and so on.

(/ )

Supplier Homepage Product Patient Monitor and Doppler Detector My-C039 12 Lead ECG+NIBP+SpO2+Pr+Resp+Temp+Printer+Etco2+IBP ICU Patient Monitor

My-C039 12 Lead ECG+NIBP+SpO2+Pr+Resp+Temp+Printer+Etco2+IBP ICU Patient Monitor

FOB Price: US $1,000-1,200 / Piece
Min. Order: 1 Piece
Min. Order FOB Price
1 Piece US $1,000-1,200/ Piece
Get Latest Price
Production Capacity: 100 PCS/Month
Transport Package: Carton Box
Payment Terms: L/C, T/T, Western Union

Request a custom order and have something just for you!

Send Customized Request
Basic Info
  • Model NO.: MY-C039
  • Type: Patient/Fetal/Maternal Monitor
  • Dimension: 365 X 150 X 340 (Mm)
  • Recording Width: 48mm
  • Trademark: MAYA
  • Origin: China
  • Classification: Physiological Functions of Diagnosis and Monitoring Equipment
  • Medical Devices Reg./Record No.: Ssyjxzz20062210132
  • Net Weight: 4.0kg
  • Paper Speed: 25mm/S, 50mm/S
  • Specification: CE, ISO
Product Description
MY-C039 12 lead ECG+NIBP+SpO2+PR+Resp+Temp+Printer+ETCO2+IBP
ICU Patient Monitor


MY-C039 features in high cost-performance, high safety and low power consumption. To ensure monitoring quality and reduce operation risk, it adopts full-isolation(floating), ECG defibrillation-proof protection, ECG anti-high frequency surgical unit and NIBP dual-overpressure protection.

Features:
Display mode: 15'' color TFT LCD with high-resolution.
Be applicable for adult, pediatric and neonatal for all-round monitoring.
Operation interface with Chinese and English.
Waveform and data can be set optionally.
Storage of 72-hour trend data, and review of 40-second holographic waveform.
Full-lead ECG display.
7-lead ECG displays in a screen, and ECG waveforms series display.
Anti-high frequency surgical unit, defibrillation-proof(requirement for special leads).
Adopt digital SpO2 technology, which has strong anti-interference and anti-weak filling capability.
Function of NIBP review, storage for up to 400 NIBP data.
Network: Connecting with central station, other Bed observation and software updating.
Network connection mode: Wireless and wired.
Built-in rechargeable battery for uninterrupted monitoring.
Standard parameters of ECG, RESP, NIBP, SpO2, TEMP and PR.
Built-in printer(optional).
Standard parameters of ECG, RESP, NIBP, SpO2, and TEMP.
IBP, CO2 and thermal printer are optional.

Performance:

ECG
Lead Mode 3-lead and 5-lead are optional
Lead Selection I, II, III, AVR, avL, avF, V
Wave 5-lead: 2 channels
3-lead: 1channel
Gain × 0.25, × 0.5, × 1, × 2

HR Measuring and Alarm Range
Range 15 ~ 300 bpm
Accuracy ± 1% or ± 1bpm, which is greater
Alarm Accuracy ± 2bpm
Resolution 1 bpm

ST Segment Monitoring
Measuring and Alarm Range -0.6 mV~ + 0.8 mV

ARR
ARR Detecting Type ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC
Alarm Available
Review Available
Scan Speed for ECG Waveform 12.5mm/s, 25mm/s, 50mm/s

Respiration (RESP)
Method R-F(RA-LL) Impedance
Measuring and Alarm Range for Resp. Rate:
Adult 0~120rpm
Pediatric and Neonatal 0~150rpm
Resolution 1 rpm
Measuring Accuracy 2 rpm
Apnea Alarm 10 ~ 40 S

SpO2
Measuring and Alarm Range 0 ~ 100 %
Resolution 1 %
Accuracy:
70% ~ 100% ± 2%
0% ~ 69% unspecified
PR
Measuring and Alarm Range 0~250bpm
Resolution 1bpm
Measuring Accuracy ± 2bpm or ± 2%, which is greater
Alarm Accuracy ± 2bpm
NIBP
Method Oscillometry
Mode Manual, Auto, continuous
Measuring Interval in AUTO Mode 1 /2 /3 /4 /5/ 10/ 15/ 30 /60 /90 Min
Measuring Period in Continuous Mode 5 Min
Measuring and Alarm Range 10~270mmHg
Resolution 1 mmHg
Accuracy ± 10% or ± 8mmHg, which is greater
Over-pressure Protection:
Adult Mode 315± 10 mmHg
Pediatric Mode 265± 10 mmHg
Neonatal Mode 155± 10 mmHg
TEMP
Channel dual-channel
Measuring and Alarm Range 0 ~ 50C
Resolution 0.1C
Accuracy ± 0.1C
Optional

EtCO2
Method Sidestream or Mainstream Measuring Range for CO2 0~150mmHg
Resolution for CO2:
0.1 mm Hg 0 to 69 mm Hg
0.25 mm Hg 70 to 150 mm Hg
Accuracy for CO2:
0 40 mm Hg ± 2 mm Hg
41 70 mm Hg ± 5%
71 100 mm Hg ± 8%
101 150 mm Hg ± 10%
Respiration Rate>80BPM ± 12%
AwRR Range 2~150 rpm
AwRR Accuracy ± 1BPM
Apnea Alarm Available

IBP
Channel dual-channel
Label ART, PA, CVP, RAP, LAP, ICP, P1, P2
Measuring and Alarm Range -50~350 mm Hg
Resolution 1 mm Hg
Accuracy ± 2% or 1mm Hg, which is greater
Send your message to this supplier
Avatar
Miss Susie
*From:
To: Miss Susie
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now

Find Similar Products By Category