My-E011 Cheapest Price for Medical Anesthesia Machine
Main body High-strength engineering plastic rack, light,
Beautiful and corrosion resistant
Scope of application adult, child
Gas source O2: 0.27~0.55MPa N2O: 0.27~0.55MPa
Flow meter O2: 0.1~1.0L/min 1.1~10L/min
N20: 0.1~1.0L/min 1.1~10L/min
Air: 0.1~10L/min
O2, N2O linkage and N2O Stopper
When using nitrous oxide, oxygen concentration>25%;
When the oxygen pressure<0.2Kpa, the flow of nitrous oxide would be cut off.
Flow rate of rapid oxygen supply 25~75L/min
Low oxygen pressure alarm There will be sound alarm when the oxygen pressure < 0.2MPa
N20
0.1~1.0L/min 1.1~10L/min
Air
0.1~10L/min
Vaporizer it has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0~5 vol%. Among Halothane, Enflurane, Isoflurane and Sevoflurane, two can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available.
Respiratory circuit working mode: All-close, semi-close, semi-open
APL: 0.5~7 kPa
Respiratory Bellows bellows for adults, tidal volume range: 0~1500 ml
Bellows for children, tidal volume range: 0~300 ml
Anyone can be chosen for application as
Required by the customer
Ventilator
Display mode High-definition 8.4" TFT color LCD screen display
Ventilation mode IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH(IMV, VCV, PCV)
Ventilation parameters Tidal volume 50~1500 ml (20~1500 ml)
Rate 1~99 bpm
SIMV rate 2~20 bpm
I: E 2: 1~1: 8(4: 1~1: 8)
Inspiratory trigger pressure -1.0~2.0 kPa
PEEP 0~2.0 kPa
Pressure range 1.0~6.0 kPa
SIGH 1.5 times the inspiratory time
(70-150 adjustable)
Inspiratory platform 0~50%
Parameters for ventilation monitoring tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, average pressure, pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure, Inspiratory platform
(autonomous inspiratory rate, lung compliance, airway resistance, flow-volume loop, pressure-volume loop)
Monitoring of oxygen concentration 21%~100%
Security Alarm System
Oxygen Concentration alarm upper limit setting range 21%~100%
Low limit setting range 10%~80%
Airway pressure alarm upper limit setting range 0.3~6.0 kPa
Low limit setting range 0.2~5.0 kPa
Per-minute ventilation volume alarm upper limit setting range 3.0~30L/min
Low limit setting range 1.0~10L/min
Sustained high-pressure alarm it will give alarm when stress have consistently
Been higher than 2.5 kPa
Suffocation alarm it will give sound and light alarm if there
Is no tidal volume input for 15 seconds
Power alarm
Power AC 220V± 10% 50Hz± 1Hz
(UPS, Storage battery)
(Mechanical Arms)
(End-expiratory CO2 monitoring)