Product Description
Hospital medical equipment single or double channel cheap syringe pump |
Single Channel
Injection Mode:Constant Rate, Time/Volume, Drug Weight, Micro, Sequential, Drug Library Mode
Syringe Size 5ml,10ml,20ml,30ml,50/60ml,auto recognition
Flow Rate 5ml:0.10ml/h-60.00ml/h 10ml:0.10ml/h-300.00ml/h
20ml:0.10ml/h-400.00ml/h 30ml:0.10ml/h-600.00ml/h
50/60ml:0.10ml/h-1200.00ml/h
Increment 0.01ml/h
Preset Time 1s~99hr59min59s
Parameter Changed during Injection Support change of VTBI, Flow rate during injection
Accuracy ≤±2%(±1% mechanical accuracy included)
Preset Volume(VTBI) 0.10-9999.99ml & empty
Accumulated Volume 0.00-9999.99ml
Purge Rate 5ml:30-60ml/h 10ml:150-300ml/h
20ml:200-400ml/h; 30ml:300-600ml/
50ml:600-1200ml/h
Bolus Rate 5ml:0.10ml/h-60.00ml/h 10ml:0.10ml/h-300.00ml/h
20ml:0.10ml/h-400.00ml/h 30ml:0.10ml/h-600.00ml/h
50/60ml:0.10ml/h-1200.00ml/h
KVO 0.10-5.0ml/h adjustable
Occlusion Pressure 8 level adjustable,20Kpa-140Kpa,dynamically pressure value display.
Alarm Finished, near finished, near empty, empty, occlusion, syringe disconnected, no operation, parameter error, syringe size error, low battery, battery lost, battery ran out, AC power lost, abnormal injection, communication error
Syringe Management Preset 20 syringe brands, can add or edit the brand, accept all brands after calibration
Display 3.5 " TFT color LCD(HF-610);
5 " TFT color LCD(HF-620);
10 level brightness adjustable
Power Supply AC Power:100V-240V,50/60Hz,<=25W
Communication Port USB,RJ45 for future use, Ethernet Port
Battery Rechargeable Li-ion battery, DC11.1V,3000mAh;operating time: ≥7h @5ml/h
Alarm Voice Standard Medical Alarm Voice with Mute function, 8 levels adjustable.
Event Log 10000 event log, can be transmitted to PC with USB
Software Upgrade Upgrade the software with USB
Operation Environment +5ºC- +40ºC,Relative humidity:20%-90%;Atmospheric pressure:70 - 106 Kpa
Classification Class II, Type CF;IPX4
Dimension Size:290mm×155 mm×225mm,Weight:3.5kg