Customization: | Available |
---|---|
Classification: | Imaging Diagnostic Equipment |
Type: | X Ray Equipment |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Description | Technology Parameters | |
Power Supply | Voltage | A.C.380V / 220V (±10%) |
Frequency | 50Hz ± 1Hz | |
Capacity | 20KVA | |
Power Supply Resistance | A.C.380V≤1.0Ω; A.C.220V≤0.3Ω | |
Maximum tube voltage | 125KV | |
Maximum tube currency | 200mA | |
Fluoroscopy | Tube voltage | 40~110KV |
Tube currency | 0.5~5mA | |
Radiograph | Tube voltage | 40~125KV |
Tube currency | Small Focus: 25mA,50mA,100mA; Big Focus:150,200mA | |
Spot film | 100mA (Small Focus) | |
Time | 0.04~6.3s Divided into 23 grades from 0.04~6.3s | |
High Voltage Generator | Capacity | 15KW |
X-ray Tube | Model | XD 51-20,40 / 125 |
Focus | Small Focus 1×1mm2 Big Focus 2×2mm2 | |
Rotated Speed of anode | 2800r / min | |
Diagnostic Table | Scope of rotation | 90°~ 0°~-5° |
Mobile beam filtration device | Grid density N=28 | |
Focus Distance f o>1000mm | ||
Grid Ratio r=8 | ||
Distance from surface to film | ≤60mm | |
Stomach & Intestine Radiograph Device | Spot film device specification(Stomach & Intestine Radiograph) | 127×178mm( 5× 7) binary picture |
203×254mm( 8×10) binary picture | ||
254×305mm(10×12) binary picture | ||
Moving lengthways (along ground railway) | 1500mm | |
Moving crosswise (left & right) | 740mm | |
Moving up and down | 700mm ~ 1900mm | |
Way of locking | Electromagnetism locking | |
Beam Limited Device | Manually single automatic multi-optional | |
Table for radiograph | Length×width×height | 2000×760×680(mm) |
Movable scope of the bed | Lengthways: ±450mm; Crossways: ±110 | |
Pillar moves lengthways | 1000mm | |
Distance from surface to film | ≤60mm | |
Bulb tube moves up and down | 1100mm~1760mm | |
Radiography item | 10×12 11×14 12×15 |