MY-B026 Microplate Reader
1. INTRODUCTION
1.1 Characters
The Microplate Reader is a semiautomated benchtop instrument focused on clinical immunity-test in hospitals or epidemic preventing stations. It has good characters such as high sensitivity, high accuracy,
high speed of detection and the easiness of operation etc. Display is a liquid crystal one with alphanumeric and
graphic options. Parameters can be checked on the display without using difficult command sequences.
1.2 Main Functions
The instrument is compatible with registered diagnostic EIA kit, primarily used in hospitals or some detection
related to clinical immunology.
2 Function and Technical Specifications
2.1Fundamental parameters
1) Range of working wavelength :400nm-800nm;
2) Range of detection:0.000A -4.000A;
3) Resolving power: 0.001 A;
4) Filters when leaving factory: 405nm, 450nm, 492nm, 630 nm, with the maximum array of 10 filters between 400nm-800nm simultaneously.
5) Reading speed: single wave 5s, dual wave 10s;
6) Dimension: 420×325×210 mm(l×h×w)
7) Weight: 10 kg
2.2Technical Specifications
(1) Characters of the filters
Table 1: Characters of the Filters
Accuracy of wavelength (nm) |
Half-width(nm) |
±3.0 |
≤12 |
(2) The errors of absorbance accuracy are as follows:
Table 2: Absorbance Accuracy
Absorbance (A) |
Accuracy Error |
0.000-2.000 |
±0.01 |
>2.000-3.000 |
±0.05 |
>3.000-4.000 |
±0.15 |
(4) The Linearity Error are as follows:
Table 3: Linearity of the Instrument
Absorbance (A) |
Error (%) |
0.000-2.000 |
±1.0 |
>2.000-3.000 |
±2.0 |
>3.000-4.000 |
±5.0% |
(5) Reproducibility Error: ≤0.5%;
(6) Stability Error: ≤±0.005A/10min;
2.3 Main function
1) Provided with power on self-checking. Error data will display on LCD monitor if something wrong with the photoelectric system.
2)Filter choosing-mode: automatically
3) Checking mode: single wave/dual wave detection; single aperture /dual aperture detection
4) Imput mode:key in touch screen;
5) Imput format: imput capital letter/lowercase letter/number at random;
6)Item parameters: Provided with function of set/save/ inquiry 120 detection parameters , to be called in later usage.
7)Defection of calibration: with the function of calibration-defection,which can save/modify 120 parameters of calibration and curves.
8)Detecting method:absorbance/qualitative/quantitative detection;
Qualitative test: single critical value checking; dual-critical value ,threshold level testing,gray-area setting;
Quantitative test:single point,parabolic interpolation, linear regression, semi-logarithm regression, logarithm regression, logit-log , enzyme kinetics, PCR testing;
9)Saving function:can save the checking results of 100 pieces of plates to be called in later usage.
10)Output mode:
Absorbance:display/printout originally absorbable value;
Qualitative checking:display/printout absorbable value, critical value, S/CO, qualitative results;
Quantitative checking:display/printout absorbable value, density, standard reference value,quantitative results.
11) QC function:inquiry/printout the QC data and chart;
12) Communication:equipped with RS-232 interface.
13) With self-monitoring function(real-time)on each testing light-source signal channel.
14) With dormancy function on light-source lamp.
2.4 Work Conditions
1) Environmental temperature: 10~30°C;
2) Relative humidity: ≤70%;
3)Power Supply : 220×(1±10%)V; 50×(1±2%)Hz; Fuse φ5×20 2A;Input power:100VA
4) Keep away from strongly electromagnetic field and avoid direct irradiation by strong light;
5) Dustproof and shockproof on worksite;no aggressive gas; pollution level:II
6) Power supply must be firmly grounded. Transient over voltage level:II.
7) Preheating time: 10min
8)Don't use instrument in damp environment